==Helley Smith sampler (HS)==
+
This article is a summary of sub-section 5.5.1.3 of the [[Manual Sediment Transport Measurements in Rivers, Estuaries and Coastal Seas]] <ref>Rijn, L. C. van (1986). ''Manual sediment transport measurements''. Delft, The Netherlands: Delft Hydraulics Laboratory</ref>. This article describes how a Helley-Smith (HS) sampler can be used to measure the [[bed load]] transport.
 +
 
 +
==Introduction==
 +
[[Image:H5513figure1.jpg|thumb|300px|right|Figure 1: Helley-Smith sampler]]
 +
The Helley-Smith [[bed-load]] sampler is a modified version of the BTMA-sampler (Helley and Smith, 1971<ref>Helley, E.J. and Smith, W., 1971. ''Development and Calibration of a Pressure-Difference Bed Load Sampler''. U.S. Geological Survey Open File Report, Washington,  USA.</ref>) (see also [[bed load transportmeter Arnhem (BTMA)]]. The Helley-Smith sampler consists of a nozzle,  sample bag and frame  (Figure 1). The sampler has a square entrance nozzle (0.076 x 0.076 m) and a sample bag constructed of 250 um-mesh polyester. Several different versions of the sampler have been used for various field conditions. Larger nozzles are generally used to sample larger sediment sizes and heavier samplers become necessary as deeper and faster rivers are sampled. An important advantage of the Helley-Smith sampler is the extensive calibration (based on about  10,000 samples) and its simple operation.
  
The Helley-Smith bed-load sampler is a modified version of the BTMA-sampler ('''Helley and Smith, 1971'''). The Helley-Smith sampler consists of a nozzle,  sample bag and frame  (Figure 1). The sampler has a square entrance nozzle (0.076 x 0.076 m) and a sample bag constructed of 250 um-mesh polyester. Several different versions of the sampler have been used for various field conditions. Larger nozzles are generally used to sample larger sediment sizes and heavier samplers become necessary as deeper and faster rivers are sampled. An important advantage of the Helley-Smith sampler is the extensive calibration (based on about  10,000 samples) and its simple operation.
 
 
The bed-load transport (in kg/sm) can be determined as:
 
The bed-load transport (in kg/sm) can be determined as:
  
S<sub>b</sub> =k s (1-p) V/(bT)    or    S<sub>b</sub>= G/(bT)
+
S<sub>b</sub> =k s (1-p) V/(bT)    or    S<sub>b</sub>= G/(bT)
  
in which: k= calibration factor, p= porosity factor (= 0.4), s= density of sediment particles (= 2650 kg/m<sup>3</sup>), V= immersed volume of sediment catch (m<sup>3</sup>), G= dry mass of sediment catch (kg), b= width of  intake opening=0.0762 m), T = sampling period (s)
+
in which: k= calibration factor, p= porosity factor (= 0.4), s= density of sediment particles (= 2650 kg/m<sup>3</sup>), V= immersed volume of sediment catch (m<sup>3</sup>), G= dry mass of sediment catch (kg), b= width of  intake opening=0.0762 m), T = sampling period (s).
  
Figures 2 and 3 present calibration  curves  relating  the  sediment catches  and  the actual transport  rates for  various  size  fractions. The actual  transport  rate  has  been  assumed  to  be represented  by simultaneous  measurements  with  a  conveyer  belt  system  just  downstream of the sampling position of the Helley-Smith sampler  ('''Emmett,  1980''').
+
==Calibration==
The  conveyer-belt  trap consists of a concrete  slot (width  = 0.4 m, depth = 0.6  m) in  the channel bed,  orthogonal  to the flow direction. Along the  bottom of  the concrete slot  passes  an endless  belt of  rubber (width = 0.3 m). Sediment falling into the open slot  drops on the moving belt, and is carried  laterally to  the  riverbank where it is scraped off the belt, weighted and returned to the river flow (depth = 1.2 m, width = 15 m,  discharge = 20 m(sup>3</sup>/s).
+
[[Image:H5513figure2.jpg|thumb|300px|right|Figure 2: Calibration curves of Helley-Smith sampler]]
Based on Figures 2 and 3, '''Emmet  (1980)''' concluded that  the Helley-Smith Sampler  has  an efficiency of    100%  (calibration factor, k = 1) for particle sizes in  the  range of  0.5  mm  to 16 mm.  For  particles  in the range  of  0.25 to 0.5  mm  the efficiency is found  to  be  about  175%.
+
Figure presents calibration  curves  relating  the  sediment catches  and  the actual transport  rates for  various  size  fractions. The actual  transport  rate  has  been  assumed  to  be represented  by simultaneous  measurements  with  a  conveyer  belt  system  just  downstream of the sampling position of the Helley-Smith sampler  (Emmett,  1980<ref name="emm">Emmett,  W.W., 1980. ''A Field Calibration of  the Sediment Trapping Characteristics of the Helley-Smith Bed Load Sampler''.  Geological Survey Professional Paper 1139, Washington, USA.</ref>).
(resulting in a calibration factor k= 0.5) which is  assumed to be caused by the collection of suspended sediment particles.
+
The  conveyer-belt  trap consists of a concrete  slot (width  = 0.4 m, depth = 0.6  m) in  the channel bed,  orthogonal  to the flow direction. Along the  bottom of  the concrete slot  passes  an endless  belt of  rubber (width = 0.3 m). Sediment falling into the open slot  drops on the moving belt, and is carried  laterally to  the  riverbank where it is scraped off the belt, weighted and returned to the river flow (depth = 1.2 m, width = 15 m,  discharge = 20 m<sup>3</sup>/s.
 +
Based on Figure 2, Emmet  (1980<ref name="emm"/>)  concluded that  the Helley-Smith Sampler  has  an efficiency of    100%  (calibration factor, k = 1) for particle sizes in  the  range of  0.5  mm  to 16 mm.  For  particles  in the range  of  0.25 to 0.5  mm  the efficiency is found  to  be  about  175%. (resulting in a calibration factor k= 0.5) which is  assumed to be caused by the collection of suspended sediment particles.
 
For particles  in the range of  16 to 32 mm the efficiency is found to be about  70% (resulting  in  a  calibration factor k = 1.5)  which is  assumed to  be  caused  by  the  paucity  of  large  particles moving as  bed load  in the flow.
 
For particles  in the range of  16 to 32 mm the efficiency is found to be about  70% (resulting  in  a  calibration factor k = 1.5)  which is  assumed to  be  caused  by  the  paucity  of  large  particles moving as  bed load  in the flow.
  
Line 19: Line 23:
 
k= 1.5, for particles in the range of 16 to 32 mm.
 
k= 1.5, for particles in the range of 16 to 32 mm.
  
More information of the calibration of bed load samplers is given by '''Hubbell et al., 1985'''.
+
More information of the calibration of bed load samplers is given by Hubbell et al., 1985<ref name="hubb">Hubbell, D.W., Stevens, H.H., Skinner, J.V.  and Beverage, J.P., 1985. New Approach to Calibrating Bed Load Samplers.  ''Journal  of Hydraulic Engineering'', Vol.111, No.4.</ref>.
 
 
The accuracy of the measured bed-load transport is strongly dependent on the accuracy of the calibration factor, the number of measurements and the sampling procedure (skilled personel).  
 
Assuming ideal sampling, at least 20 samples must be collected at each location to obtain a bed-load transport rate with a standard deviation (error) of about 20% ('''De Vries, 1973'''). In practice the sampling error will be considerably larger (say 100%) particularly due to the sampling procedure (see also '''Hubbell et al. 1985''').
 
  
 +
The accuracy of the measured [[bed-load]] transport is strongly dependent on the accuracy of the calibration factor, the number of measurements and the sampling procedure (skilled personnel). Assuming ideal sampling, at least 20 samples must be collected at each location to obtain a [[bed-load]] transport rate with a standard deviation (error) of about 20% (De Vries, 1973<ref>De Vries, M.,  1973. ''On Measuring Discharge and Sediment Transport in River Flow''. Delft Hydraulics Laboratory, Publication No. 106, The Netherlands.</ref>). In practice the sampling error will be considerably larger (say 100%) particularly due to the sampling procedure (see also Hubbell  et al. 1985<ref name="hubb"/>).
  
 +
==See also==
 +
===Summaries of the manual===
 +
* [[Manual Sediment Transport Measurements in Rivers, Estuaries and Coastal Seas]]
 +
* Chapter 1: [[Introduction, problems and approaches in sediment transport measurements]]
 +
* Chapter 2:  [[Definitions, processes and models in morphology]]
 +
* Chapter 3: [[Principles, statistics and errors of measuring sediment transport]]
 +
* Chapter 4: [[Computation of sediment transport and presentation of results]]
 +
* Chapter 5: [[Measuring instruments for sediment transport]]
 +
* Chapter 6: [[Measuring instruments for particle size and fall velocity]]
 +
* Chapter 7: [[Measuring instruments for bed material sampling]]
 +
* Chapter 8: [[Laboratory and in situ analysis of samples]]
 +
* Chapter 9: [[In situ measurement of wet bulk density]]
 +
* Chapter 10: [[Instruments for bed level detection]]
 +
* Chapter 11: [[Argus video]]
 +
* Chapter 12: [[Measuring instruments for fluid velocity, pressure and wave height]]
  
==References==
+
===Other internal links===
 +
* [[Bottle and trap samplers]]
 +
* [[Bed load transportmeter Arnhem (BTMA)]]
 +
* [[Delft Nile bed load and suspended load sampler (DNS)]]
 +
* [[Sand transport]]
  
'''Delft Hydraulics, 1996'''. Test measurements of Helley Smith and modified Helly Smith (in Dutch). Report Q 2141, Delft, The Netherlands
 
 
'''Delft Hydraulics, 1997'''. Calibration and comparison of Helley Smith. Report Q 2345,Delft, The Netherlands
 
 
'''De Vries, M.,  1973'''. On Measuring Discharge and Sediment Transport in River Flow. Delft Hydraulics Laboratory, Publication No. 106, The Netherlands
 
 
'''Emmett,  W.W., 1980'''. A Field Calibration of  the Sediment Trapping Characteristics of the Helley-Smith Bed Load Sampler.  Geological Survey Professional Paper 1139, Washington, USA
 
 
'''Helley, E.J. and Smith, W., 1971'''. Development and Calibration of a Pressure-Difference Bed Load Sampler. U.S. Geological Survey Open File Report, Washington,  USA
 
 
'''Hubbell, D.W., Stevens, H.H., Skinner, J.V.  and Beverage, J.P., 1985'''. New Approach to Calibrating Bed Load Samplers.  Journal  of Hydraulic Engineering,  Vol.111,  No.4
 
  
  
 +
===Further reading===
 +
* Delft Hydraulics, 1996. ''Test measurements of Helley Smith and modified Helly Smith'' (in Dutch). Report Q 2141, Delft, The Netherlands.
 +
* Delft Hydraulics, 1997. ''Calibration and comparison of Helley Smith''. Report Q 2345,Delft, The Netherlands.
  
 +
==References==
 
<references/>
 
<references/>
 
==See also==
 
 
===Other contributions of Leo van Rijn===
 
 
====articles with parts of the manual====
 
*[[Manual Sediment Transport Measurements in Rivers, Estuaries and Coastal Seas]]
 
 
*[[INTRODUCTION, PROBLEMS AND APPROACHES IN SEDIMENT TRANSPORT MEASUREMENTS]]
 
*[[DEFINITIONS, PROCESSES AND MODELS IN MORPHOLOGY]]
 
*[[PRINCIPLES, STATISTICS AND ERRORS OF MEASURING SEDIMENT TRANSPORT]]
 
*[[COMPUTATION OF SEDIMENT TRANSPORT AND PRESENTATION OF RESULTS]]
 
*[[MEASURING INSTRUMENTS FOR SEDIMENT TRANSPORT]]
 
*[[MEASURING INSTRUMENTS FOR PARTICLE SIZE AND FALL VELOCITY]]
 
*[[MEASURING INSTRUMENTS FOR BED MATERIAL SAMPLING]]
 
*[[LABORATORY AND IN-SITU ANALYSIS OF SAMPLES]]
 
*[[IN-SITU MEASUREMENT OF WET BULK DENSITY]]
 
*[[INSTRUMENTS FOR BED LEVEL DETECTION]]
 
*[[ARGUS VIDEO]]
 
*[[MEASURING  INSTRUMENTS FOR FLUID VELOCITY, PRESSURE AND WAVE HEIGHT]]
 
 
 
==External links==
 
 
==Crediting the authors==
 
 
  
 
{{author  
 
{{author  
 
|AuthorID=13226  
 
|AuthorID=13226  
|AuthorName= Rijn, Leo van}}
+
|AuthorFullName= Rijn, Leo van
 
+
|AuthorName=Leovanrijn}}
  
 
{{author  
 
{{author  
 
|AuthorID=12969  
 
|AuthorID=12969  
|AuthorName= Roberti, Hans}}
+
|AuthorFullName= Roberti, Hans
 +
|AuthorName=Robertihans}}
 +
 
 +
[[Category:Coastal and marine observation and monitoring]]
 +
[[Category:Observation of physical parameters]]

Latest revision as of 13:39, 19 August 2020

This article is a summary of sub-section 5.5.1.3 of the Manual Sediment Transport Measurements in Rivers, Estuaries and Coastal Seas [1]. This article describes how a Helley-Smith (HS) sampler can be used to measure the bed load transport.

Introduction

Figure 1: Helley-Smith sampler

The Helley-Smith bed-load sampler is a modified version of the BTMA-sampler (Helley and Smith, 1971[2]) (see also bed load transportmeter Arnhem (BTMA). The Helley-Smith sampler consists of a nozzle, sample bag and frame (Figure 1). The sampler has a square entrance nozzle (0.076 x 0.076 m) and a sample bag constructed of 250 um-mesh polyester. Several different versions of the sampler have been used for various field conditions. Larger nozzles are generally used to sample larger sediment sizes and heavier samplers become necessary as deeper and faster rivers are sampled. An important advantage of the Helley-Smith sampler is the extensive calibration (based on about 10,000 samples) and its simple operation.

The bed-load transport (in kg/sm) can be determined as:

Sb =k s (1-p) V/(bT)    or     Sb= G/(bT)

in which: k= calibration factor, p= porosity factor (= 0.4), s= density of sediment particles (= 2650 kg/m3), V= immersed volume of sediment catch (m3), G= dry mass of sediment catch (kg), b= width of intake opening=0.0762 m), T = sampling period (s).

Calibration

Figure 2: Calibration curves of Helley-Smith sampler

Figure 2 presents calibration curves relating the sediment catches and the actual transport rates for various size fractions. The actual transport rate has been assumed to be represented by simultaneous measurements with a conveyer belt system just downstream of the sampling position of the Helley-Smith sampler (Emmett, 1980[3]). The conveyer-belt trap consists of a concrete slot (width = 0.4 m, depth = 0.6 m) in the channel bed, orthogonal to the flow direction. Along the bottom of the concrete slot passes an endless belt of rubber (width = 0.3 m). Sediment falling into the open slot drops on the moving belt, and is carried laterally to the riverbank where it is scraped off the belt, weighted and returned to the river flow (depth = 1.2 m, width = 15 m, discharge = 20 m3/s. Based on Figure 2, Emmet (1980[3]) concluded that the Helley-Smith Sampler has an efficiency of 100% (calibration factor, k = 1) for particle sizes in the range of 0.5 mm to 16 mm. For particles in the range of 0.25 to 0.5 mm the efficiency is found to be about 175%. (resulting in a calibration factor k= 0.5) which is assumed to be caused by the collection of suspended sediment particles. For particles in the range of 16 to 32 mm the efficiency is found to be about 70% (resulting in a calibration factor k = 1.5) which is assumed to be caused by the paucity of large particles moving as bed load in the flow.

Summarizing: k= 0.5, for particles in the range of 0.25 to 0.5 mm, k= 1.0, for particles in the range of 0.5 to 16 mm, k= 1.5, for particles in the range of 16 to 32 mm.

More information of the calibration of bed load samplers is given by Hubbell et al., 1985[4].

The accuracy of the measured bed-load transport is strongly dependent on the accuracy of the calibration factor, the number of measurements and the sampling procedure (skilled personnel). Assuming ideal sampling, at least 20 samples must be collected at each location to obtain a bed-load transport rate with a standard deviation (error) of about 20% (De Vries, 1973[5]). In practice the sampling error will be considerably larger (say 100%) particularly due to the sampling procedure (see also Hubbell et al. 1985[4]).

See also

Summaries of the manual

Other internal links


Further reading

  • Delft Hydraulics, 1996. Test measurements of Helley Smith and modified Helly Smith (in Dutch). Report Q 2141, Delft, The Netherlands.
  • Delft Hydraulics, 1997. Calibration and comparison of Helley Smith. Report Q 2345,Delft, The Netherlands.

References

  1. Rijn, L. C. van (1986). Manual sediment transport measurements. Delft, The Netherlands: Delft Hydraulics Laboratory
  2. Helley, E.J. and Smith, W., 1971. Development and Calibration of a Pressure-Difference Bed Load Sampler. U.S. Geological Survey Open File Report, Washington, USA.
  3. 3.0 3.1 Emmett, W.W., 1980. A Field Calibration of the Sediment Trapping Characteristics of the Helley-Smith Bed Load Sampler. Geological Survey Professional Paper 1139, Washington, USA.
  4. 4.0 4.1 Hubbell, D.W., Stevens, H.H., Skinner, J.V. and Beverage, J.P., 1985. New Approach to Calibrating Bed Load Samplers. Journal of Hydraulic Engineering, Vol.111, No.4.
  5. De Vries, M., 1973. On Measuring Discharge and Sediment Transport in River Flow. Delft Hydraulics Laboratory, Publication No. 106, The Netherlands.
The main author of this article is Rijn, Leo van
Please note that others may also have edited the contents of this article.

Citation: Rijn, Leo van (2020): Helley-Smith sampler (HS). Available from http://www.coastalwiki.org/wiki/Helley-Smith_sampler_(HS) [accessed on 25-11-2024]


The main author of this article is Roberti, Hans
Please note that others may also have edited the contents of this article.

Citation: Roberti, Hans (2020): Helley-Smith sampler (HS). Available from http://www.coastalwiki.org/wiki/Helley-Smith_sampler_(HS) [accessed on 25-11-2024]